Spatially resolved measurements of ballistic and total transmission in microscale tissue samples from 450 nm to 1624 nm - Tout INSP Accéder directement au contenu
Article Dans Une Revue Biomedical optics express Année : 2021

Spatially resolved measurements of ballistic and total transmission in microscale tissue samples from 450 nm to 1624 nm

Résumé

We built a simple and versatile setup to measure tissue ballistic and total transmission with customizable wavelength range, spatial resolution, and sample sizes. We performed ballistic transmission and total transmission measurements of overlying structures from biological samples ex vivo. We obtained spatially resolved transmission maps to reveal transmission heterogeneity from five microscale tissue samples: Danionella skin, mouse skull bone, mosquito cuticle, wasp cuticle, and rat dura over a wide spectral range from 450 nm to 1624 nm at a spatial resolution of ∼25 µm for ballistic transmission measurements and ∼50 µm for total transmission measurements. We expect our method can be straightforwardly applied to measuring the transmission of other samples. The measurement results will be valuable for multiphoton microscopy. The total transmission of a sample is important for the collection of multiphoton excited fluorescence and the assessment of laser-induced sample heating. The ballistic transmission determines the excitation power at the focus and hence the fluorescence signal generation. Therefore, knowledge of ballistic transmission, total transmission, and transmission heterogeneity of overlying structures of animals and organs are essential to determine the optimal excitation wavelength and fluorophores for non-invasive multiphoton microscopy.
Fichier principal
Vignette du fichier
boe-13-1-438.pdf (24.55 Mo) Télécharger le fichier
Origine : Publication financée par une institution

Dates et versions

hal-03519194 , version 1 (10-01-2022)

Identifiants

Citer

Aaron T Mok, Jamien Shea, Chunyan Wu, Fei Xia, Rose Tatarsky, et al.. Spatially resolved measurements of ballistic and total transmission in microscale tissue samples from 450 nm to 1624 nm. Biomedical optics express, 2021, 13 (1), pp.438. ⟨10.1364/boe.441844⟩. ⟨hal-03519194⟩
14 Consultations
40 Téléchargements

Altmetric

Partager

Gmail Facebook X LinkedIn More